VERTICAL DEFLECTION
(2 IDENTICAL CHANNELS) |
|
Bandwidth* and Rise Time (at all deflection
factors from 50 Ω terminated source) |
|
-15°C to +40°C |
+40°C to +55°C |
|
465 |
Dc to 100 MHz, 3.5 ns |
85 MHz, 4.1 ns |
|
475 |
Dc to 200 MHz, 1.8 ns |
175 MHz, 2.0 ns |
|
475A |
Dc to 250 MHz, 1.4 ns |
250 MHz, 1.4 ns |
|
*Measured at -3 dB, Bandwidth may be limited
to approx 20 MHz by bandwidth limit switch. |
|
Lower -3 dB point, ac coupling 1X probe: 10
Hz or less. 10X probe: 1 Hz or less. |
|
Deflection Factor at BW |
465: 5 mV/div to 5 V/div
475: 2 mV/div to 5 V/div
475A: 5 mV/div to 10 V/div |
|
1-2-5 sequence, accurate ± 3%. Uncalibrated,
continuously variable between steps and to
at least 12.5 V/div (465/475) to at least 25
V/div (475A). In cascade mode sensitivity
is approx 1 mV/div (465); approx 400 µV/div
(475);
and approx 2.5 mV/div (475A).
Cascaded bandwidth is at least 50 MHz
(465/475/475A)
when signal out is terminated in 50 Ω. |
|
Display Modes |
Ch 1; Ch 2 (normal and inverted), alternate,
chopped (465-approx 250 kHz rate,
475/475A-approx 1 MHz rate),
added; X-Y (Ch 1-X, Ch 2-Y) |
|
CMRR |
Common-mode rejection ratio at least 20 dB
at 20 MHz
(50 MHz for 475/475A)
for common-mode signals of 6 div or less |
|
Automatic Scale Factor |
Probe tip deflection factors for 1X or 10X
coded probes are automatically indicated by
two readout lights behind the knob skirts.
All lights are off when the channel is not
displayed. Ground reference display
selectable at probe (when dc coupled). |
|
Input R and C |
1 MΩ ± 2% paralleled by approx 20 pF. |
|
Max Input Voltage |
Dc Coupled |
250 V (dc + peak ac)
500 V (p-p ac at 1 kHz or less) |
|
Ac Coupled |
500 V (dc + peak ac)
500 V (p-p ac at 1 kHz or less) |
|
Delay Line |
Permits viewing leading edge of displayed
waveform |
|
Probe Power (475/475A only) |
Connectors provide correct voltages for two
optional P6201 FET Probes |
|
HORIZONTAL DEFLECTION |
|
465 |
|
Time Base A |
0.05 µs/div to 0.5 s/div (1-2-5 sequence).
X10 mag extends max sweep rate to 5 ns/div. |
|
Time Base B |
0.05 µs/div to 50 ms/div (1-2-5 sequence).
X10 mag extends max sweep rate to 5 ns/div. |
|
475/475A |
|
Time Base A and B |
0.01 µs/div to 0.5 s/div (1-2-5 sequence).
X10 mag extends max sweep rate to 1 ns/div. |
|
Variable Time Control |
Time Base A (465/475/475A) provides
continuously variable uncalibrated sweep
rates between steps and to at least 1.25
s/div. Warning light indicates uncalibrated
setting. |
|
Time Base A and B Accuracy, full 10 cm
|
Unmagnified |
+20°C to +30°C |
-15°C to +55°C |
|
465 |
475/475A |
465 |
475/475A |
|
± 2% |
± 1% |
± 3% |
± 2% |
|
Magnified |
± 3% |
± 2% |
± 4% |
± 3% |
|
Horizontal Display Modes |
A, mixed sweep, A intensified, B delayed. B
ends A for increased intensity in the
delayed mode. |
|
Calibrated Mixed Sweep |
Displays A sweep for period determined by
delay-time position control, then displays B
sweep for remainder of horizontal sweep. |
|
CALIBRATED SWEEP DELAY |
|
Delay
Time
Range |
465:
0.2 to X10 delay time/div settings of 200/ns
to 0.5 s (minimum delay time is 200 ns).
475/475A:
0 to X10 delay time/div settings of 50 ns to
0.5 s (minimum delay time is 50 ns). |
|
Differential Time Measurement Accuracy |
Delay Time Setting |
+15°C to 35°C |
|
Over one or more major dial divisions |
± 1% |
|
Less than one major dial division |
± 0.01 major dial division |
|
Jitter |
1 part or less in 50,000 (0.002%) of 10X the
A sweep time/div setting. 1 part in 20,000
when operating from 50 Hz line. |
|
TRIGGERING A AND B |
|
A Trigger Modes |
Normal (sweep runs when triggered).
Automatic (sweep free-runs in the absence of
a triggering signal and for signals below 30
Hz). Single sweep (sweep runs one time on
the first triggering event after the reset
selector is pressed). Lights indicate when
sweep is triggered and when single sweep is
ready. |
|
A Trigger Hold off |
Adjustable control permits a stable
presentation of repetitive complex
waveforms. |
|
B Trigger Modes |
B runs after delay time (starts
automatically at the end of delay time) and
B triggerable after delay time (runs when
triggered). The B (delayed) sweep runs
once, in each of these modes, following the
A sweep delay time. |
|
Time Base A and B Sensitivity and Coupling |
|
Coupling |
465 |
475 |
475A |
|
To 25 MHz |
At 100 MHz |
To 40 MHz |
At 200 MHz |
To 40 MHz |
At 250 MHz |
|
DC |
Internal
|
0.3 div deflection |
1.5 div deflection |
0.3 div deflection |
1.5 div deflection |
0.3 div deflection |
2.0 div deflection |
|
External |
50 mV |
150 mV |
50 mV |
250 mV |
50 mV |
250 mV |
|
External +10 |
500 mV |
1.5 V |
500 mV |
2.5 V |
500 mV |
2.5 V |
|
Ac |
Requirements increase below 60 Hz |
|
Ac Lf Reject |
Requirements increase below 50 kHz |
|
Ac Hf Reject |
Requirements increase below 60 Hz and above
50 kHz |
|
465 Jitter |
0.5 ns or less at 100 MHz and 5 ns/div |
|
475 Jitter |
0.2 ns or less at 200 MHz and 1 ns/div |
|
475A Jitter |
0.2 ns or less at 250 MHz and 1 ns/div |
|
A Trigger View |
A spring-loaded pushbutton overrides other
vertical controls and displays the external
signal used for A sweep triggering. This
provides quick verification of the signal
and time comparison between a vertical
signal and their trigger signal. |
|
Level and Slope |
Internal, permits selection of triggering at
any point on the positive or negative slope
of the displayed waveform. Level adjustment
through at least ± 2 V in external, through
at least ± 20 V in external divided by 10. |
|
A Sources |
Norm, Ch 1, Ch 2, line, external, and
external divided by 10 |
|
B Sources |
Starts after delay, norm, Ch 1, Ch 2, and
external |
|
External Inputs |
R and C approx 1 MΩ paralleled by approx 20
pF. 250 V (dc + peak ac) max input. |
|
X-Y OPERATION |
|
465 |
|
Full-sensitivity X-Y (Ch1 Horiz, Ch2 Vert) |
5 mV/div to 5 V/div, accurate ± 4%.
Bandwidth is dc to at least 4 MHz. Phase
difference between amplifiers is 3° or less
from dc to 50 kHz. |
|
475, 475A |
|
Full-sensitivity X-Y (Ch1 Horiz, Ch2 Vert) |
2 mV/div to 5 V/div (475), 5 mV to 10 V/div
(475A) accurate ± 3%. Bandwidth is dc to at
least 3 MHz. Phase difference between
amplifiers is 1° or less from dc to 1 MHz. |
|
DISPLAY |
|
CRT |
8 X 10 cm display. Horizontal and vertical
centerlines further marked in 0.2 cm
increments. P31 phosphor standard; P11
option. 18 kV accelerating potential. |
|
Graticule |
internal, nonparallax; variable edge
lighting; markings for measurement of rise
time |
|
Beam Finder |
Compresses trace to within graticule area
for ease in determining the location of an
offscreen signal. A pre-set intensity level
provides a constant brightness. |
|
Z-Axis Input |
Dc coupled, positive-going signal decreases
intensity; 5 V p-p signal causes noticeable
modulation at normal intensity; dc to 50
MHz. |
|
ENVIRONMENTAL CAPABILITIES |
|
Ambient Temperature |
Operating:
-15°C to +55°C.
Nonoperating: -55°C to +75°C.
Filtered forced air ventilation is provided. |
|
Altitude |
Operating:
to 15,000 ft; max allowable ambient
temperature decreased by 1°C/1000 ft from
5000 to 15,000 ft.
Nonoperating: to 50,000 ft. |
|
Vibration |
Operating:
15 minutes along each of the three axes, .06
cm (0.025 in) p-p displacement (4 g's at 55
Hz) 10 to 55 to 10 Hz in 1 minute cycles. |
|
Humidity |
Operating and nonoperating:
5 cycles (120 hours) to 95% relative
humidity referenced to MIL-E-16400F (par
4.5.9 through 4.5.9.5.1, class 4). |
|
Shock |
Operating and nonoperating:
30 g's 1/2 sine, 11 ms duration, 2 shocks
per axis in each direction for a total of 12
shocks. |
|
OTHER CHARACTERISTICS |
|
Amplitude Calibrator |
Output Voltage |
0.3 V |
1%
0°C to +40°C |
|
Output Current |
30 mA |
|